Salbutamol ELISA Diagnostic Kit |
 |

heart,
Urine,
muscle,
Milk
and feed could be as low as 0.25 ppb. If the sample has been purified with
column, the sensitivity can even reach 0.05 ppb.
|
Specificity |
Cross reactivity of the corresponding substances. |
|
Cross-reactivity |
Compound |
Cross-Reactivity |
|
Salbutamol |
100% |
|
Terbutaline |
90% |
|
Clenbuterol |
70% |
|
Isoproterenol |
29% |
|
Sensitivity of Samples |
Matrix |
Sensitivity (ppb) |
|
Serum, Muscle, Organ (with C18 column
) |
0.05 |
|
Serum |
0.25 |
|
Muscle |
0.5 |
|
Organ (liver, heart) |
0.5 |
|
Feed |
2.5 |
|
Recovery rate |
Sample
(spike
SB in ppb) |
Recovery rate |
|
Muscle
(Spike
1.0 ppb) |
90~120% |
|
Organ (Spike
2.5ppb) |
90~130% |
|
Serum (Spike
1.25 ppb) |
90~110% |
|
Feed
(Spike 2.0ppb) |
80~115% |
|
Operation Benefit |
Operation time : 90 mins |
|
|
|
 |
|
|
|
|
|
 |

Taiwan Advance Bio-Pharm Inc.
All Rights Reserved.
12F., No.23,
Ln. 169, Kangning St., Xizhi Dist., New Taipei City 22180, Taiwan (R.O.C.)
TEL: +886-2-2695-9766 FAX:
+886-2-2695-8518 E-mail:
service@abc-residue.com

|
|
 |